Collimator-Based Tracking with an Add-On Multileaf Collimator : Modification of a commercial collimator system for realtime applications (Bestmeddiss)

個数:

Collimator-Based Tracking with an Add-On Multileaf Collimator : Modification of a commercial collimator system for realtime applications (Bestmeddiss)

  • オンデマンド(OD/POD)版です。キャンセルは承れません。
  • ≪洋書のご注文について≫ 「海外取次在庫あり」「国内在庫僅少」および「国内仕入れ先からお取り寄せいたします」表示の商品でもクリスマス前(12/20~12/25)および年末年始までにお届けできないことがございます。あらかじめご了承ください。

  • 【入荷遅延について】
    世界情勢の影響により、海外からお取り寄せとなる洋書・洋古書の入荷が、表示している標準的な納期よりも遅延する場合がございます。
    おそれいりますが、あらかじめご了承くださいますようお願い申し上げます。
  • ◆画像の表紙や帯等は実物とは異なる場合があります。
  • ◆ウェブストアでの洋書販売価格は、弊社店舗等での販売価格とは異なります。
    また、洋書販売価格は、ご注文確定時点での日本円価格となります。
    ご注文確定後に、同じ洋書の販売価格が変動しても、それは反映されません。
  • 製本 Paperback:紙装版/ペーパーバック版/ページ数 53 p.
  • 商品コード 9783658106577

Full Description

In this thesis, a tracking system was developed by modifying an add-on collimator, the Siemens Moduleaf, for realtime applications in radiotherapy. As the add-on collimator works almost completely autonomously of the linear accelerator (LinAc), no modifications to the latter were necessary. The adaptations to the Moduleaf were mainly software-based. In order to reduce the complexity of the system, outdated electronic parts were replaced with newer components where practical.

Verification was performed by measuring the latency of the system as well as the impact on applied dose to a predefined target volume, moving in the leaf's travel direction. Latency measurements in software were accomplished by comparing the target and current positions of the leaves. For dose measurements, a Gafchromic EBT2 film was placed beneath the target 4D phantom, in between solid water plates, and moved alongside with it.

Based on the results, a tracking-capable add-on collimator seems to be a useful tool for reducing the margins for the treatment of small, slow-moving targets.

Radiotherapy is one of the most important methods used for the treatment of cancer. Irradiating a moving target is also one of the most challenging tasks to accomplish in modern radiotherapy.

Contents

1 Introduction.- 2 Literature Overview.- 3 Hypothesis & Goals.- 4 Materials & Methods.- 5 Results.- 6 Discussion.- 7 Conclusions.